5.2 External Connections

The external connections of the converter are shown in Figure 5.1. See 5.4 "Wiring Method" to
connect the cables correctly
Terminal block
IV wire 5.5mm
or more
2
Class D grounding
(Ground resistance
100 or less)
Thick steel conduit
Flow rate signal cable
(2-core shielded cabtyre cable)
Connected detector
Excitation cable
(3-core cabtyre cable)
5.1 External Wiring Connection Diagram
Output cable (CVV-S)
Input/Output cable (CVV-S)
- 21 -
6F8A0917
Power cable (CVV)
Power supply
Current output
(4 20mADC)
Digital input 2 (option)
Digital input 1 (option)
(20 30VDC)
Digital output 4 (option)
Digital output 3 (option)
Digital output 2 (option)
Digital output 1