Toshiba L5232 Instruction Manual page 85

Electromagnetic flowmeter converter
Table of Contents

Advertisement

Changing the current output value used when an alarm occurs
The following is an example to show how to change the current output value, which is used when
an alarm occurs, from 4.0mA (2: 4.0mA) to 24.0mA or more (4: OVER 24.0mA).
Switch operation
Display example
D3: 4-20 ALM.OUT
D3: 4-20 ALM.OUT
D3: 4-20 ALM.OUT
D3: 4-20 ALM.OUT
D3: 4-20 ALM.OUT
Select D3: 4−20 ALM. OUT from the setting item selection
menu.
The currently set alarm output value used when an alarm
2:4.0mA
occurs appears.
The cursor appears.
2:4.0mA
Press
4:OVER 24.0mA
The display blinks and the set value is displayed for
confirmation.
Press
4:OVER 24.0mA
disappears and the display returns to the condition that the
setting value is displayed.
Returns to the menu of setting item selection.
- 84 -
Description
to select 4: OVER 24.0mA.
again to write the data. Then the cursor