Analog Rx Signal Path - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-9-19-2
Analog Receive Signal Path and Analog Rx Path (cont'd)

7-9-19-2-1Analog Rx Signal Path

...
Analog Rx Tests (High, Medium and Low)
Test signal generated on GFI, and injected into GRX.
Tests frequency, amplitude, SNR, THD.
Reports failures at channel level.
Test both GRX Boards and A/D Converter on DRX Boards.
PROPRIETARY TO GE
. 7
EV
Figure 7-184 Analog Rx Signal Path
GRX64 Ch
GRX128 Ch
192
aTGC
aTGC
Chapter 7 Diagnostics/Troubleshooting
DRX 64
DRX 64
DRX 64
Nathan
A/D
0
Nathan
1
Nathan
2
Nathan
A/D
3
GDIF
LOGIQ E9 P
ROPRIETARY
GFI
FDEM
SDEM
BEP
ATGC
T_sig
S
M
ERVICE
ANUAL
7 - 279

Advertisement

Table of Contents
loading

Table of Contents