Close Up Of Transmit Channel - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-9-19-7
Visual Inspection of Board (cont'd)
Transmit Channel
4) Transmit Transformer
5) Local Power Resistors
When Visual Inspection Reveals Physical Damage
Replace the GTX Board that shows evidence of burn damage. Then run the following diagnostics:
-
-
Failure of one of these diagnostics, following the replacement of the defective GTX Board, will typically
mean damage to one of the Front Planes. Swap locations of the Front Planes and re-run the tests to
see if the failure follows the Front Plane. In rare occurrences damage has occurred to the GRX Board.
Damage to the GRX Board may show as a bright or dark spot in the near field image. See:
Figure 7-192 "Check in near Field Image" on page 7-288.
If these diagnostics pass check image quality in the near field:
-
PROPRIETARY TO GE
. 7
EV
Figure 7-191 Close up of Transmit Channel
Run T/R Channel diagnostic
Run Analog Rx diagnostics
Install and check image quality on all probes that may have been used during the time period
that the system began to show the transformer failure issue.
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL
7 - 287

Advertisement

Table of Contents
loading

Table of Contents