Spectrometer Autocalibration

Spectrometer Autocalibration
The MB3000 series spectrometer is designed to automatically perform an
autocalibration every 2 weeks. The autocalibration will start between two
measurements but never during a measurement. When the autocalibration is started,
the MB3000 spectrometer will be inoperative for about 2 minutes. The Status LED
will flash. When autocalibration has finished, the spectrometer resumes operation.
Spectrometer is inoperative during autocalibration.

MB3000 Series Spectrometers Calibration

A spectral calibration must be performed in these conditions:
To perform spectral calibration, follow this procedure:
Spectrometer must be under power for at least 4 hours before performing spectral
calibration.
During calibration, spectrometer does not acquire any data.
STEP 9
STEP 10
STEP 11
After step 3, the spectrometer automatically performs the calibration which lasts
approximately three minutes.
When the calibration is completed, the spectrometer resumes operation.
16
After installation, once the temperature of the spectrometer has stabilized i.e.
when the spectrometer has been under power for four hours.
Anytime the spectrometer is powered ON after being powered OFF for more
than one hour.
Set the ON/OFF power switch of the spectrometer power supply to OFF.
Wait 30 seconds.
Set the ON/OFF power switch of the spectrometer power supply to ON.
Section 3 Unpacking and Installing
AA003700-01 rev. G. 2.0