LED troubleshooting table
.
LED
Power
Status
40
State
Possible cause
OFF
Power switch not ON
Power cable not connected
No current at Power outlet
Internal problems
GREEN
Normal operation
RED
No detector connected
blinking
GREEN
with Red
Operation Error
short blinks
RED
Internal Error
OFF
System is loading
System is initializing or
performing calibration.
GREEN
The initialization sequence is
Blinking
performed at startup or after a
Table- 9- 1. LED Troubleshooting table
perturbation.
Section 9 Troubleshooting
Solution
Turn power switch ON. Refer to
Connector Location
Make sure the power cable is
connected in the Spectrometer and in
the wall outlet. Refer to
Location
on page 26
Test outlet for voltage.
Contact ABB. Refer to the back cover
of this document
Replace or install detector module.
Refer to
Detector Module
Replacement
on page 29.
Check software health monitoring.
Check software health monitoring.
If after 1 min. LED is still OFF, check
software health monitoring
If this sequence last longer than 5
minutes, it is probably caused by a
low laser or white light level
Contact ABB. Refer to the back cover
of this document.
AA003700-01 rev. G. 2.0
on page 26.
Connector