Table 11-33 Calomat 7: Climatic Conditions - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis, rack-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
11.3 Technical specifications of analyzer modules

Table 11-33 CALOMAT 7: Climatic conditions

Climatic conditions
Storage and transport
Max. permissible ambient temperature
surrounding the basic device during op‐
eration
Ambient humidity (rel. humidity) during
storage, transportation or operation
Restriction for installing an ULTRAMAT 7 analyzer module: 5 to 45 °C
1)
Avoid condensation on the installed components!
2)
Table 11-34 CALOMAT 7: Gas connections
Gas connections
Connection fittings
Table 11-35 CALOMAT 7: Parts in contact with the sample gas
Materials in contact with the sample gas
Gas connection
Clamping rings and union nut (set)
Sample gas pipes
Sensor mounting block
Sensor
Gasket, contained in the sensor module Perfluorelastomere FFKM
134
1)
- 30 to 70 °C
0 to 50 °C
< 90 %
Pipe diameter 6 mm (outer diameter)
Stainless steel material no. 1.4571
Stainless steel material no. 1.4401
Stainless steel material no. 1.4404
Stainless steel material no. 1.4571
Si, SiO
N
, Au, epoxy resin, glass
x
y
Operating Instructions, 05/2015, A5E31874006-03
Rack-mounted device