Connecting

7.1

Electrical connections

7.1.1

Note on safety

Note
If auxiliary power is required, check that it corresponds with that on the nameplate and with
the test certification valid for your country.
7.1.2

Connecting the power supply

Requirements
Note
Appliance plug
An appliance plug is included in the scope of delivery. However, you can use any appliance
plug which meets the requirements of DIN EN 60320-1, ≥6 A.
Note
Connection information
The power cable must comply with the valid regulations and conditions for the place of
installation and be provided with a protective conductor at the potential of the enclosure. The
cross-section of each wire must be ≥ 1 mm
● Connect the phase conductor, neutral conductor (N) and the protective conductor (PE) to
the designated position in the plug.
● Make sure that a power disconnection device is located in the direct vicinity of the device
(for load rating, see nameplate).
● Make sure that the supply voltage matches the voltage specified on the nameplate of the
device.
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03
.
2
7
83