Application planning
4.1 Notes on applications planning
OXYMAT 7
Due to its functional principles, the analyzer module emits magnetic stray fields. Therefore, do
not operate magnetically sensitive devices in the immediate vicinity of the analyzer module.
Depending on the sensitivity, a distance of up to 50 cm from the analyzer module is required.
Note
Impact-free installation location
Building vibrations, in particular in the frequency range from 6 to 15 Hz, can influence the
measured signal.
Select an installation location that is free of vibration.
CALOMAT 7
The CALOMAT 7 is not affected by building vibrations and shocks.
Introduction of flammable gases or gas mixtures
Only introduce flammable gases or gas mixtures when they are non-explosive.
58
WARNING
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03