HP 53150A Series Programming Manual page 105

Microwave frequency counter
Table of Contents

Advertisement

Chapter 4 Command Reference
:DISPlay Subsystem
:DISPlay Subsystem
This subsystem controls the selection and presentation of textual
information on the Counter's display. This information includes
measurement results. :DISPlay is independent of, and does not modify,
how data is returned to the controller.
See the section titled "Programming the Counter to Display Results" on
page 3-46 of this guide.
:DISPlay[:WINDow]:BACKground[:STATe] . . . <Boolean>
COMMAND
Turns the display backlight ON or OFF.
QUERY RESPONSE
COMMENTS
RELATED
FRONT-PANEL
KEYS
Programming Guide
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: ON
Turning the display backlight OFF conserves battery power (if the
Counter has the Battery option).
If the Counter is operating from battery power (Battery option
required), the display backlight is automatically turned off after a
period of inactivity. When activity is subsequently detected (a
measurable signal is applied, a front-panel key is pressed, or a setting
is changed by a command sent over the HP-IB or the RS-232
interface), the display backlight is restored to its previous state (i.e., if
the backlight was set to OFF, it remains off; if the backlight was set to
ON, it is turned on).
On/Off (Shift + Clear)
4-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

53152a series53151a series

Table of Contents