HP 53150A Series Programming Manual page 147

Microwave frequency counter
Table of Contents

Advertisement

Chapter 4 Command Reference
Common Commands
*ESE
(Standard Event Status Enable Command)
*ESE?
(Standard Event Status Enable Query)
*ESE <NRf> | <non-decimal numeric>
COMMAND
*ESE?
Sets or queries the Standard Event Status Enable Register, shown in
Figure 4-3.
The parameter and query response value, when rounded to an integer
value and expressed in base 2 (binary), represents the bit values of the
Standard Event Status Enable Register. The value of unused bits is zero
when queried and ignored when set.
This register is used to enable a single or inclusive OR group of Standard
Event Status Register events to be summarized in the Status Byte
Register (bit 5).
Figure 4-3. The Standard Event Status Enable Register
See the section titled "Standard Event Status Register" on page 3-30 of
this guide for a detailed description of the Standard Event Status
Register.
Programming Guide
Bits
( 1 )
Operation Complete
Not Used
( 4 )
Query Error
( 8 )
Device Dependent Error
( 16 )
Execution Error
( 32 )
Command Error
Not Used
( 128 )
Power On
0
1
2
3
4
5
6
7
4-47