Chapter 4 Command Reference
[:SENSe]:FUNCtion Subtree
[:SENSe]:FUNCtion[:ON] . . . <sensor_function>[,<sensor_function>]
COMMAND
Selects the sensor functions to be sensed by the Counter.
The supported <sensor_function> strings are:
•
QUERY RESPONSE
•
•
•
COMMENTS
•
•
Various
RELATED
FRONT-PANEL
KEYS
4-28
"[XNONe:]FREQuency [1 | 2]"
"[XNONe:]POWer [2]"
The query form of this command returns a comma-separated list of
functions that are ON.
The string omits default nodes (XNONe) and uses short-form
mnemonics. If the channel specifier(s) are set to default value(s), no
channel specifier is returned in response. If the channel specifier(s)
are not set to default value(s), they are returned in the response with
a single space separating the first channel specifier from the function
name.
For example:
–
"FREQ" is returned for frequency on Channel 1.
–
"FREQ 2" is returned for frequency on Channel 2.
The only functions that can be turned ON simultaneously are "POW
2" and "FREQ 2".
*RST: "FREQ 1"
This command can be used to turn individual function(s) ON without
affecting other functions.
If the optional channel specification is omitted from the
<sensor_function>, a default channel selection is made. For Frequency
and Power, the default is Channel 2.