DANGER!
Mains voltage
When wiring the power supply line, ensure that the double or reinforced insulation between the
electric circuits and the secondary circuit remains intact.
NOTICE
Insulation
The additional insulation must be suitable for a test voltage of 1500 VAC. The thickness of the
insulation must be at least 0.4 mm (e.g. insulating tubing, type BI 85 from Bierther GmbH).
The additional insulation of the power cables (mains connection, alarm and signalling relays) can be implemented as follows:
- Terminals (plug-type connectors)
1
2
- Flexible insulating tubing for the power cables
3
- Connecting cable
METPOINT
BDL compact
®
3
3
Installation and operating manual