Home  Settings  Sensor settings  A1  right arrow (2. page)  Gas constant
The remaining text fields can be labelled and configured in the same manner. For details, see chapter 11.2.2.7 Labelling and
configuring text fields!
Home  Settings  Sensor settings  A1  right arrow (2. page)
Note:
After confirming with OK, the field labels change back to black and the settings are applied.
Caution:
Reference temperature and reference pressure (factory settings 20°C, 1000 hPa): All volume flow (m³/h) and 
consumption (m³) values shown on the display refer to 20°C and 1000 hPa (according to ISO 1217). Alternatively, enter 
0°C and 1013 hPa (=standard cubic metre according to DIN 1343) as the reference values. Do not enter the operating
pressure or the operating temperature as the reference values!
METPOINT
BDL compact
®
Installation and operating manual
Preselection of matching Units.
If a text field is displayed with red text, the 
respective values (e.g. Diameter or Name) have been
changed.