6-2-3
LCD Adjustment Procedure
Table 6-1
Dark Room for Radiology/Cardiology
Dim Room for Radiology/Cardiology
Bright Room
Factory Settings
6-2-3-1
Brightness
To reduce the brightness, press the left arrow button. A popup box with the current brightness
setting will be displayed on the screen.
To increase the brightness, press the right button.
Push the mode button once to remove the popup bow from the screen and save the current values.
6-2-3-2
Contrast
Use the recommended LCD settings. The contrast is adjusted by the LOGIQ E9 software, when
selecting scanning modes and doing scanning adjustments. Changing the monitor's contrast settings in
one mode will influence the picture quality in other modes.
. 2
EV
Recommended LCD Settings
Room Condition
Chapter 6 Service Adjustments
Contrast
85
90
100
100
LOGIQ E9 S
M
ERVICE
ANUAL
Brightness
35
55
100
60
6 - 3