GE
D
5535208-100, R
IRECTION
7-5-3
Diagnostics Window Overview
1.) Instructions Frame
-
2.) Status Frame
-
-
3.) Loop Count
-
4.) Execute Button
-
-
-
5.) Progress Indicator
-
6.) Short Text Message
-
7.) Status Frame Background Color
-
-
-
-
7 - 22
. 2
EV
Displays either test-specific text or the default instructions.
Initially displays the last known status for a selected diagnostic. Once the diagnostic starts,
the frame displays the "current" status of all test results. Also see:
on Diagnostic Window" on page 7-24
The Status Frame also contains the user interface elements used for Diagnostic Control and
Operator Feedback.
This is an editable text field that only accepts numeric values of 4 digits or less. When the
switch is configured as an "execute" switch and pressed, the loop count field will be queried
to determine the number of times to execute the diagnostic.
This switch has two modes - each with appropriate text:
Execute - to start the diagnostic
Abort - to stop a diagnostic
Displays a graphical progress indicator for the user.
Displays brief messages about the test's progress during execution.
Initially gray, the Status Frame background color changes upon completion of a diagnostic to
indicate completion status.
Code Status Fail = Red
Code Status Pass = Green
Code Status Abort = Yellow
Section 7-5 - Common Service Desktop
for more current status data reporting.
LOGIQ E9 S
M
ERVICE
ANUAL
Figure 7-26 "Details Link