Keyboard
See:
7-5-3 "Diagnostics Window Overview" on page 7-22
1.) Click the Execute button on the Keyboard window. The PC-Doctor Keyboard Test opens.
2.) Select your keyboard type from the drop-down menu. Usually the default choice is the correct one.
3.) Click the Start button.
4.) Press each key on the system's keyboard once and make sure the corresponding keys on-screen
are removed from view.
-
-
-
5.) The test status is displayed in the Current Status portion of the window.
7 - 84
. 2
EV
Click the Pass button if all the keys are removed from the PC-Doctor Keyboard Test.
Click the Fail button if any key is not removed from the on-screen keyboard. Failed keys may
be damaged and you may have to have your keyboard repaired or replaced.
Click the Abort button to exit the test.
Section 7-5 - Common Service Desktop
LOGIQ E9 S
for a description of this window's features.
M
ERVICE
ANUAL