Download Print this page

GE LOGIQ E9 Service Manual page 159

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
4-2-3-1
Complete Power Down
Table 4-1
1.
Power down the system and disconnect the
mains power cable from the wall outlet.
Press once on the ON/OFF button on the
Operator Panel to display the System -
Exit menu. When you switch off the unit,
the system performs an automatic
shutdown sequence.
2.
Select Shutdown to do a complete power
down of the LOGIQ E9.
The Back End Processor will first turn off
the scanner activity and print the message
"Please wait - Shutdown in progress" in the
LCD display on the Operator Panel.
Next, it starts to shut down itself. The time
to shutdown the LOGIQ E9, including the
Back End Processor, may vary from 10
seconds up to approximately 1 minute.
The last thing that shuts down is the light on
the Operator Panel, indicating that you can
continue with the next step.
NOTE: Sleep Mode only available in
LOGIQ E9 running R2.x.x and R3.x.x
software.
Be sure to wait with the next step until the system has finished its shut-down. Failing to do so may
destroy data on the Hard Disk Drive, making the system fail later.
Switch off the Mains Power Circuit Breaker, located on the rear of the LOGIQ E9. This will cut power
distribution within the unit.
Unplug the unit.
5
EV
Power Down the LOGIQ E9
Steps
NOTICE
Chapter 4 Functional Checks
LOGIQ E9 Service Manual
Corresponding Graphic
Press once on the ON/OFF button
4 - 9

Advertisement

loading