Download Print this page

GE LOGIQ E9 Service Manual page 542

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
Table 8-89
2.
3.
8 - 178
5
EV
Main Power Supply (PS) removal
Steps
If the 4D MC option is installed, disconnect
the USB Cable from J28 on the BEP. The
4D MC will also have to be transferred to
the replacement PS. See Section
"4D Motor Controller (4D MC)
replacement" on page
8-186.
Ease the PS away from the fang bracket.
Pull/slide the entire PS backwards until the
connectors to the Card Cage/Backplane
are unseated.
Lift the PS away.
Section 8-10 - Main Power Supply parts replacement
8-10-4
LOGIQ E9 Service Manual
Corresponding Graphic