GE LOGIQ E9 Service Manual page 700

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
10-6-3
Leakage Current Limits (cont'd)
Table 10-12 Equipment Type and Test Definitions
Applied Parts
(AP)
Type BF
Type CF
Sink
Leakage
10 - 12
5
EV
Parts or accessories that contact the patient to perform their function. For ultrasound
equipment, this includes transducers and ECG leads.
Body Floating or non-conductive ultrasound
probes which are marked with the 'man in box'
BF symbol. this includes all transducers.
Cardiac Floating or non-conductive
intraoperative probes for direct cardiac contact
and isolated ECG connections so marked with
the 'heart in box' CF symbol.
The current resulting from the application of mains voltage to the applied part. This test
is required test for Type CF applied parts.
Section 10-6 - Electrical Safety Tests
LOGIQ E9 Service Manual