8-5-3
Top Cover replacement
8-5-3-1
Manpower
One person, 15 minutes.
8-5-3-2
Tools
For tools needed, refer to:
8-5-3-3
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system
2.) Disconnect the mains power cable from the wall outlet.
3.) Disconnect all Probes and External I/O Cabling.
4.) Remove the Side Covers.
Follow these links if you need more information:
4-2-4 "Power shut down" on page 4-7.
8-5-2 "Side Covers replacement" on page 8-27.
8-5-3-4
Top Cover removal
1.) Remove the two screws that secure the Top Cover.
2.) Remove the Top Cover and gently pull back and up.
5
EV
8-2-5 "Tools needed for servicing LOGIQ E9" on page
Figure 8-20 Top Cover Screw placement (seen from above)
Chapter 8 Replacement procedures