6
Warnings and troubleshooting
报 报 报 报 警 警 警 警 及 及 及 及 故 故 故 故 障 障 障 障
Check the setting for the Q2 limit.
Replace the polisher module if it is depleted.
检查进水水质(电导率是否超标);
检查 Q2 的极限值设定是否正确;
如果抛光柱失效,请予以更换。
报 报 报 报 警 警 警 警
Warning
The temperature has exceeded the limit:
Check the setting for the temperature limit (35 ° C) .
Check whether the system is heating up.
Check the temperature of the feed water
Open the dispenser and discharge water
Set interval mode
检查温度的极限设定值是否正确(35 ° C) ;
检查纯水机本身是否过热;
检查进水的水温;
打开水龙头排掉一些水;
将 UV 灯设定成周期运行模式。
报 报 报 报 警 警 警 警
Warning
high:RO 压力太高
RO pressure too
Check feed water pressure.
Check RO module for blockages and replace if necessary.
Check restriction for blockages.
Inform customer service.
Adjust pressure regulator.
检查进水压力;
检查 RO 膜是否被污堵,如必要的话就换掉;
检查限流器是否被污堵;
通知客户服务中心;
校正压力表。
After eliminating the cause of this error message, you must
disconnect the system from the electricity supply for at least 30
seconds to delete the message!!!
报 报 报 报 警 警 警 警
Warning
更换 UV 灯管
Replace UV lamp
UV lamp has been in operation for maximum time.
Replace UV lamp 4.1.
UV4.1 灯管的使用时间已经达到设定值,需要更换。
LaboStar 7 TWF DI and UV
Ultra pure water system
Catalog no. 2231 & 2232
水温超过设定值
13061776706@163.com