Siemens SIREMOBIL Iso-C Maintenance Instruction page 23

Table of Contents

Advertisement

General Information
Technical safety checks, test certificate
of the check
2.3 Are undamaged accessories being
used?
2.4 System radiation protection devices
present and undamaged? (No on-site radi-
ation shield)
2.5 Are the required operator documents
complete, present and legible?
2.6 Are all warning labels in place and rec-
ognizable?
2.7 Are the operating symbols, light indica-
tors and button labels OK?
3. Electrical checks
3.1 Measurement of protective conductor
resistance
3.2 Leakage current or equivalent leakage
current measurement
Siemens AG
Medical Solutions
SPR2-230.101.02.06.02
07.05
Maintenance instructions for the
check/comments
3. Mechanical safety, checks:
SIM I.I. laser light localizer mechanics
SIM I.I. laser light localizer function
SIM Laser light localizer mechanics
SIM Laser light localizer function
SIM Navigation system
SIM DICOM bridge mounting
SIM Codonics printer mounting
3. Electrical safety, checks:
SIE Fluoroscopic timer
SIE Audible warning signal
SIE Compulsory radiation switch-off
SIE Check of the radiation release
switches
SIE Dose rate
SIE Radiation indicators
SIE Iris collimator
SIE Area dose product measuring unit
Operating instructions must be available.
Ask the customer for additional required
operator documents and check them for
completeness, availability, and legibility.
3. Mechanical safety, checks:
SIM Warning signs
SIM ID labels
3. Electrical safety, check:
SIE Check of the operating functions
n.a.
3. Electrical safety, check:
SIE Protective conductor test
5.1 Final work steps, check:
SIE Protective conductor test
n.a.
Page 23 of 46