Siemens SIREMOBIL Iso-C Maintenance Instruction page 43

Table of Contents

Advertisement

Maintenance, Operating Value / Functional Inspection 43
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible. The previously indicated monitor display of position and angle of rotation must
coincide with the actual position of the slot diaphragm.
Select I.I. zoom format.
Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible.
The previously indicated monitor display of position and angle of rotation must coincide
with the actual position of the slot diaphragm.
Open the slot diaphragm completely and rotate it again.
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible.
The previously displayed monitor display must match the position and angle of rotation
of the actual slot diaphragm position.
NOTE
Siemens AG
Medical Solutions
Perform this check with the lowest possible kV values so that the
front edges of the slot diaphragm plate are effectively visualized.
SPR2-230.101.02.06.02
07.05
Page 43 of 46
CS PS 24
SIREMOBIL Iso-C