Disconnecting The Probe - GE Vivid i User Manual

Hide thumbs Also See for Vivid i:
Table of Contents

Advertisement

Probes
To select a probe and an application:
1.
Press
connected probes will pop up. The list will contain the name
of the probe in the main connector and the name of the CW
pencil probe connected to the small connector.
2.
Trackball to the desired probe.
An application menu for the desired probe is then listed.
3.
Trackball to the desired application
4.
Press
Make sure that the probe and application names displayed on the
screen correspond to the actual probe and application selection.
CAUTION
Check that the correct TI category is displayed (see "Thermal
Index" on page 353). TIB must be displayed when a fetal
application is selected.

Disconnecting the probe

To disconnect probes:
1.
Push the locking lever on the Vivid i all the way down.
The probes that are
2.
Remove the connector from the port.
not connected to the
3.
Ensure that the probe head is clean before placing the
unit should be
probe in its storage case.
stored in their stor-
age case.
270
on the control panel. A list of the
APPLICATION
to launch the application.
SET
Vivid i User's Manual
2378958-100 Rev. 02