Rack-Mounted Device; Device-Specific Safety Instructions - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Note
Dimensioning of the purging gas outlet
The following applies for the dimension of the minimum internal diameter and length of the
purging gas outlet:
• The permissible internal pressure in the enclosure may not be exceeded by the selected
design.
• The volume flow of purging gas during the pre-purging phase must amount to at least 60 l/
min for the selected design. For additional information about the purging gas refer to the
section "Technical specifications for operation in hazardous areas (Page 96)".
4.4

Rack-mounted device

4.4.1

Device-specific safety instructions

Risk of explosion resulting from improper installation
• Install the devices only in control cabinet/enclosure types that meet the requirements of EN/
• An air exchange of at least 5-times the volume of the enclosure per hour, but at least 164
• OXYMAT 7: Never introduce occasionally explosive gases!
NOTICE
Insufficient air circulation around the device
Device damage
The device may overheat in case of heat accumulation.
• Make sure that sufficient air circulates around the rack-mounted device.
• Do not cover the ventilation slit.
• If you install more than one rack-mounted device, maintain a clearance of one height unit
Devices in explosion-proof design
Compact Operating Instructions, 01/2022, A5E35134047-11
WARNING
IEC 60079-7 and are correspondingly approved.
l/h, is required for the introduction of combustible gases. Ensure the specified air exchange
using suitable ventilation measure.
(4.5 cm) to the next device.
Installing
4.4 Rack-mounted device
29