Siemens SIPROCESS GA700 Compact Operating Instructions page 91

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

General
Weight (only module)
Electrical safety
Max. EMC error during operation with one
and/or two OXYMAT 7/ ULTRAMAT 7 units
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
1)
Table 9-22
ULTRAMAT 7: Measuring ranges
Measuring ranges
Number of meas‐
Max. 4; parameters can be assigned freely
uring ranges
(combination with O7)
Table 9-23
ULTRAMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pressure Normal pressure (atmospheric pressure
compensation)
Pressure loss between sample gas inlet and sample gas outlet
Sample gas flow
ULTRAMAT 7 standard version
ULTRAMAT 7 high-temperature
Sample gas temperature and humidity
ULTRAMAT 7 standard version
ULTRAMAT 7 high-temperature
Nominal values in the section "Determining the technical specifications"
1)
Wall-mounted device only
2)
Devices in explosion-proof design
Compact Operating Instructions, 01/2022, A5E35134047-11
ULTRAMAT 7
Max. 5.2 kg
Component
Smallest possible measuring span
CO
CO
2
CH
4
C
H
2
4
SO
2
NO
N
O
2
NH
3
CO/NO
CO/NO
CO 0 ... 100/NO 0 ... 300 ppm
CO
/CO
2
In accordance with DIN EN 61010-1
1.0% according to EN 61326-1 (2013)
0 ... 10 ppm
0 ... 5 ppm
0 ... 50 ppm
0 ... 300 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 100 ppm
0 ... 100 ppm
500 to 1 500 hPa (absolute)
< 10 hPa at 1.5 l/min
18 ... 90 l/h (0.3 ... 1.5 l/min)
36 .. 90 l/h (0.6 ... 1.5 l/min)
0 to 50 °C (32 ... 122 °F) at relative humidity < 90%, in
reference to the respective ambient temperature; avoid
condensation in gas path (moisture content results in
measurement errors)
0 to 90 °C (32 ... 194 °F), avoid condensation in the gas path
(moisture content results in measurement errors)
Technical specifications
9.2 General technical specifications
ULTRAMAT 7 high-temperature
Max. 7.2 kg
Largest possible measuring span
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 3%
0 ... 100%
0 ... 100%
0 ... 10000 ppm
CO 0 ... 3000 ppm/NO 0 ... 3000 ppm
0 ... 100%
1, 2)
1, 2)
1)
91

Advertisement

Table of Contents
loading

Table of Contents