Calomat 7 Technical Specifications - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
9.2 General technical specifications
9.2.6

CALOMAT 7 technical specifications

Table 9-32
CALOMAT 7: General technical specifications
General
Information on the smallest span
Power consumption
Weight
Electrical safety
Max. EMC error during operation of one and/or
two CALOMAT 7 units
Table 9-33
CALOMAT 7: Measuring ranges
Measuring ranges
Number of measuring ranges
Note
The following metrological data apply to the gas pair H
measuring spans, the time behavior and the measurement behavior as well as the influencing variables.
Parameters can be assigned in the measuring
ranges
Smallest possible measuring span
Largest possible span
Smallest possible span with suppressed zero
point
Table 9-34
CALOMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pressure
Pressure drop between sample gas inlet and outlet < 50 hPa at 1.5 l/min
Sample gas flow
Sample gas temperature and humidity
Table 9-35
CALOMAT 7: Temperature of sample chamber and gas path
Temperature of sample chamber and gas path
Gas path
Standard version
94
See module nameplate
< 20 W
Approx. 3 kg
In accordance with DIN EN 61010-1
2.0% according to EN 61326-1 (2013)
Max. 4; parameters can be assigned freely
in N
. With other gas pairs, these data may differ from the possible
2
2
0.5% H
in N
2
2
100% H
in N
2
2
5% (e.g. 95 to 100%) H
in N
2
700 to 1 400 hPa (absolute)
30 to 90 l/h (0.5 to 1.5 l/min)
0 ... 70 °C (32 ... 158 °F) with relative humidity < 90%, avoid condensation
in the gas path (moisture content results in measurement errors)
5 K above ambient temperature
72 °C (162°F)
Compact Operating Instructions, 01/2022, A5E35134047-11