Requirements For Electrical Connection; Electric Connection Of Rack-Mounted Device; Connecting Signal And Ethernet Cables - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Connecting
5.2 Electrical connections
Note
Electromagnetic compatibility (EMC)
You can use this device in industrial environments and small businesses. Metal housings show
increased electromagnetic compatibility compared to high-frequency radiation. The protection
from high-frequency radiation can be increased by grounding the housing.
Note
Improvement of interference immunity
• Lay signal cables separately from cables with voltages > 60 V.
• Use cables with twisted wires.
• Install the device and lay its cables at a distance to strong electromagnetic fields.
• Use shielded lines to guarantee the full specification according to Ethernet/PROFIBUS.
Note
Unstable plug-in connections
Plug-in connections without interlock can loosen and fall out. Operation with Ex py or Ex ec can
cause an ignition.
• Secure plug-in connections without interlock mechanically, for example, with a cable tie.
5.2.4

Requirements for electrical connection

• You have ensured that a power disconnection unit belonging to the device is located in the
direct vicinity of the device.
This power disconnection device must be labelled so that it is linked exclusively with this
device.
• You have checked that the existing power supply corresponds to that specified on the rating
plate of the device.
5.2.5

Electric connection of rack-mounted device

5.2.5.1

Connecting signal and Ethernet cables

Note
Connecting signal cables
Install the signal lines separately from the power lines.
50
Devices in explosion-proof design
Compact Operating Instructions, 01/2022, A5E35134047-11