6. Work accompanying replacement of parts

Use the actions listed below as a reference when replacing the parts and units as part of the
market service.
(1) When the detector has been replaced
• Recognize the detector by infrared communication (using the service tools).
• Delete the registration of the detector which has just been replaced if necessary (using the
service tools).
• Check the firmware and FPGA versions (and update them if necessary).
• Check that booting is performed correctly.
• Proceed with calibration.
• Conduct self-diagnosis.
• Carry out the performance test.
• Check the image quality.
(2) When the X-ray interface box and PCA-XIF board have been replaced
• Change the IP address (but only when it has been changed from the default setting).
• Check the firmware versions (and update them if necessary).
• Check that booting is performed correctly.
• Check that the operations for taking and obtaining the images and for transferring the data can
be performed normally.
(3) When the AC/DC box unit, PCA-PB board and AC/DC power supply have been replaced
• Check that booting is performed correctly with the wired connection.
• Check that the operations for taking and obtaining the images and for transferring the data can
be performed normally.
(4) When the detector cable has been replaced
• Check that booting is performed correctly with the wired connection.
• Check that the operations for taking and obtaining the images and for transferring the data can
be performed normally.
4. Repair Guide
- 25 -