7.1.9 When using the detector in multiple imaging room
Users can carry a detector from one imaging room to another and use it in the other rooms only if the
service person has installed an X-ray interface box and access point in each of the rooms concerned. At
this time, bring the detector close to the infrared data communication unit, perform LINK between the
two units (detector registration), and select the settings to establish the communication circuit.
* Even when the optional wiring unit is used and connection is accomplished by wired
communication, LINK must be performed first.
* When bringing the detector from X-ray room 1 into X-ray room 2 to use it in X-ray room 2,
LINK must be performed to the infrared data communication unit in X-ray room 2, and the
settings for the detector must be changed Similarly, when returning the detector from X-ray
room 2 to X-ray room 1 to use it in X-ray room 1, LINK must be performed again to the
infrared data communication unit in that room.
* The settings are retained even after the power of the detector has been turned off. When using
the unit in the same system, LINK is not required again. If the settings are erased due to
trouble, LINK must be performed again.
X-ray room 1
Imaging room
X-ray
Access
interface box
point
Hub
Operation
room
Image Capture
Computer
2. Installation procedures
Detector
X-ray generator
- 32 -
X-ray room 2
Imaging room
X-ray interface
Access
box
point
Hub
Operation
room
Image Capture
Infrared data
communication unit
Computer
LAN inside
hospital/clinic