Canon CXDI-80C Service Manual page 15

Table of Contents

Advertisement

Item
Service life
Momentary
max. load
Power
Assumed
consumption
continuous
/heat
max. load
*1
generation
mode
Standby
Detector withstand load
operation
Control software
Photo timer
Allowable temperature of
detector panel
*1
: The target unit is a complete CXDI-80C Wireless set (including detector, X-ray interface box,
wiring unit, and status indicator). The values indicated incorporate a large margin from the
actual values or indicate attainable values.
1) Power consumption [VA]: Information for the purpose of designing the power supply
1
.
O
v
e
1
.
O
v
e
CXDI-80C
7 years
Operation time: 21,000 hours (10 hours a day,
300 days a year, 7 years)
Standby time: 61,500 hours (24 hours a day,
365 days a year, 7 years)
Approx. 70 VA or less
Approx. 20 W or less (17 kcal/h, 72 kJ/h or
less)
Approx. 15 W or less
Load applied over entire surface: 1.47 kN
(150 kgf) for 1 minute
Load applied in local areas: 0.98 kN (100 kgf),
40 mm diameter for 1 minute
There must be no problems in operation after
the loads have been applied under the above
conditions and then removed.
CXDI Control Software NE
Ver.1.40 and up supported
Cannot be incorporated
Max. +45°C (with upper limit of usable
environment estimated at +35°C, a relative
temperature of 10°C or less must be satisfied
at room temperature)
- 7 -
r
v
i
e
w
r
v
i
e
w
*2
Remarks
Ambient temperature:
+23°C ±2°C
Relative humidity:
60% ±15% RH
Atmospheric pressure:
860 to 1060 hPa
When used under the
ambient conditions
above
Momentary max.
value, excluding inrush
currents
Mean value during
continuous operation
with exposure at
15-second intervals
Sleep mode
Test conditions:
The detector shall be
placed on a rigid flat
surface with its X-ray
incidence surface face
up, and images shall be
taken with the loads on
the left brought to bear
on the effective area.
When an operating
temperature of +49°C
or more has been
detected for the
detector, a restriction
applies to the detector
where it is forcefully
set to the sleep mode
and imaging is
prohibited, and the
imaging prohibition is
released when the
temperature drops to
under +48°C.