Motorola R-2550 Programming Reference Manual page 177

Communications system analyzer
Table of Contents

Advertisement

E?
Description:
Query Syntax:
Response Syntax:
Parameters:
The E?
query returns the
Event Status Register.
Error
00
Inquiry
01
Invalid command or query mnemonic (prefix)
02
Invalid command or query mnemonic (suffix)
03
Numeric data too large
04
Numeric data too small
05
string data field too large
06
String data field too small
07
Invalid string field
08
Invalid input data
09
Invalid command or query for R-2600 mode
10
Invalid
11
Exponent out of range
Invalid
12
13
Transmission error (RS-232 only)
14
Input Buffer overflow (RS-232 o&y)
15
Empty
command
16
Output Buffer overflow
Voltmeter out of range
17
18
No input signal
19
Frequency counter measurement invalid
20
Frequency error measurement out of range
Monitor modulation measurement out of range
21
22
Input signal too high
23
Input signal too low
24
Invalid
25
Reserved
26
Thermal Power too high (LPA Option only)
27
Thermal Power too low (LPA Option Only)
28
Thermal Power Meter not zeroed (LPA Option Only)
Unable to zero Thermal Power Meter (LPA Option
29
only)
98
Error queue overflow
Error queue empty
99
E?
sl
Error Number
Description:
Type:
8 character suing
Format:
167
next error message from the error queue
Description
without
measurement
exponent
format
mantissa
format
measurement
reading
of the form "ERROR XX"
Motorola Communications Analyzer
Error
and clears the Error
command
Programming Reference