WSINGEN
Description:
Command Syntax:
Parameters:
SINAD Generate Selection
The WSINAD command selects the SINAD Generation Level.
WSINGEN nl ,n2
al
Description:
Optional:
Type:
Units:
Resolution:
Range:
Reset:
DZ.
Description:
Optional:
Type:
Units:
Resolution:
Range:
Reset:
SIN AD Generate Level Format
Yes
<Decimal Numeric Program Data>
None
NIA
0:
µV
1:
mV
2:
dBu
3:
dBm
3
SINAD Generate Level
Yes
<Decimal Numeric Program Data>
See n2 and notes
0.01 µV or0.01 mV orO.I dBu orO.I dBm
RF
110:
0.07 µV
<= SINAD Gen Level<= 707.6 µV
GEN :
22.36 µV
<= SINAD Gen Level<= 999.99 µV
GEN :
0.02 mV
<= SINAD Gen Level<= 223.87 mV
RF 1/0:
-23 dBu
<= SINAD Gen Level <= 57 dBu
GEN :
27 dBu
<= SINAD Gen Level <= 107 dBu
RFI!O:
-130dBm
<= SINADGenLevel<=-50dBm
GEN :
-80 dBm
<= SINAD Gen Level <= 0 dBm
OdBm
Notes:
The limits are dependent upon whether the RF/IO pon is selected or the GEN pon.