9.2.2 50 % Flow Rate Calibration
Using Mag-Prover's internal calibration circuit, 50% flow rate calibration can be confirmed.
9.2.3 100 % Flow Rate Calibration (Span Calibration)
Using Mag-Prover's internal calibration circuit, 100% flow rate calibration can be performed. For
calibration procedure, see the calibration procedure for 0% flow rate. (For 100% flow rate calibration,
select "CAL 100%" from the setting menu.)
9.2.4 Checking the Excitation Current
You can monitor the exciting current value.
●Checking the exciting current value
Switch operation
STEP1
Buy: www.ValinOnline.com | Phone 844-385-3099 | Email: CustomerService@valin.com
Display example
R:CAL
Select "EX DSPL" in the setting item
1 CAL
0%
selection screen.
2 CAL 50%
3 CAL 100%
4 EXC DSPL
5 EXIT
ESC
EXC DSPL
The excitation current value appears.
0.1998
Pushing
menu.
A
ESC
- 109 -
6 F 8 A 0 9 2 8
Description
ESC
returns you to the setting