5.4.2 Terminal Treatment of Cables

Follow the procedures below to treat the terminals (at the converter side) of various cables and install the
cables to the terminal block. Use appropriate cables based on the description in Section 5.1 "Cables."
Crimp a round type insulated crimp-type terminal to the end of the cables.
(1) Power cable, current output cable, and digital I/O cables
The necessary cables should be ordered from the person responsible for the installation. Strip the sheath of
each conductor as shown in Figure 5.6 and attach a crimping terminal with insulated sleeve to it. The size
of the crimping terminal is M3.5.
• Connect the power cable to terminal blocks L1 and L2.
• Connect the current output cable to terminal blocks + and -.
• Connect the digital I/O cable to terminal blocks D1, D01, D02 and COM, as required.
Figure 5.6 Terminal Treatment of Power Cable, Current Output Cable and Digital I/O cable
(2) Excitation cable
Strip the sheath from the end of each conductor as shown in Figure 5.7, attach an M3.5 crimping terminal
LF622F
with insulated sleeve, and connect it to the terminal blocks X and Y. Connect the red conductor to terminal
block E.
Buy: www.ValinOnline.com | Phone 844-385-3099 | Email: CustomerService@valin.com
25~45mm
X Black
E Red
Y White
25~45mm
Figure 5.7 Terminal Treatment of Excitation Cable
- 29 -
Crimping terminal M3.5
M3.5 crimping terminal
6 F 8 A 0 9 2 8