9. Calibration
9.1 Calibration Items
When you check or calibrate the converter or check the excitation current, you have to change the mode to
the calibration mode.
You can check or change the zero and span of the converter and the excitation current value as described
below.
However, calibration is already performed when shipped from the factory. Do not perform
change calibration unless it is specificity required.
Items
9.2.1
9.2.2
9.2.3
9.2.4
Function items
0 % flow rate calibration
50 % flow rate calibration
100 % flow rate calibration
Checking the excitation current output
− 106 −
6 F 8 A 0 9 2 7
Display example
CAL
0%
CAL 50%
CAL 100%
EXC DSPL