Switch operation
STEP6 
 
 
STEP7(=END) 
 
 
Note: The setting range of excitation current value is from 0.0000A to 0.2500A.
If you try to set an exciting current value larger than 0.2500A, an error message appears
and the setting value returns to the previous value.

8.2.3 Meter Size

Proceed as follows to check or change the meter size of the detector.
 
Switch operation
STEP1 
 
 
STEP2 
 
 
STEP3 
 
 
 
Display example
EXC CUR 
 
0.2150 
 
 
         A 
SET OK? 
ESC 
O K
EXC CUR 
 
0.2150 
 
O K
 
         A 
 
ESC 
Display example
A:DETECTOR 
1 EXC CUR 
2 SIZE 
3 EXC FREQ 
 
4 FLOW DIR 
5 EXIT 
ESC
 
SIZE 
 
 
 
        mm 
 
ESC 
SIZE 
 
  
 
        mm 
 
SIZE 
 
65 
 
        mm 
 
 
Pushing
the setting is OK or not.
If OK, push
setting, push
Pushing
ESC
and exits the setting screen.
N O
When you push
message, the set value is fixed.
Then pushing
setting menu. If you push
change the current value from 0.2150A
again.
Select "SIZE" from the setting item
selection menu.
The currently set meter size (50mm in this
example) appears.
50 
Then push
Pushing
ESC
menu.
 
The switches at the bottom change.
(
50 
At the same time, the cursor appears.
(The item indicated by the cursor is
highlighted.)
Then push
You can continue to change the setting
value.
Push
items.
: Selected item is scrolled up.
: Selected item is scrolled down.
− 56 −
6 F 8 A 0 9 2 7  
Description
shows a message to confirm
. If you need to redo the
O K
.
N O
cancels the setting operation
to "SET OK?"
O K 
brings you back to the
ESC
, you can
Description
.
returns you to the setting
are shown.)
.
 