Switch operation
STEP3 
 
 
STEP4(=END) 
 
 
Note: If you try to set a value exceeding 25.0%, HIGH OVER SPEC error appears and the

8.2.10 Damping Constant

The damping constant is used to moderate output fluctuations. (The larger the damping constant, the more
the output is averaged. But the response to an input change will be slower.) The damping constant can be
set as follows:
The damping constant is set for 0.0 sec, 0.5 sec and 1 to 60 sec (in increments of 1 second)
Note: 0.0 sec setting will work as equal to 0.1 sec damping constant.
Set 1 sec or more for normal operation.
If you set a value exceeding 60s, it is forcibly changed to 60s before data is written.
Proceed as follows to check or change the damping constant.
Shown below is an example of changing the damping constant from 2.0s to 10s.
Switch operation
STEP1 
 
 
Display example
R HYS 
 
      05.0 
 
 
         % 
 
 
 
R HYS 
 
      05.0 
 
 
         % 
SET OK? 
 
ESC 
O K
 
value returns to the previous value. Set a value once again.
Display example
D:FILTER 
1 DAMPING 
2 LIM RATE 
3 LIM TIME 
 
4 EXIT 
 
ESC
 
DAMPING 
 
      02.0 
 
 
         S 
 
 
ESC 
 
Push
desired digit and push
number of the digit.
 
When you push
to confirm whether the setting is OK or not.
If OK, push
setting, push
Pushing
ESC
and exits the setting screen.
N O
Select "DAMPING" from the setting item
selection menu.
 
 
The currently set damping constant (2.0s in
this example) appears.
Then push
 
− 72 −
6 F 8 A 0 9 2 7  
Description
to move the cursor to the
to change the
, a message appears
. If you want to redo the
O K
.
N O
cancels the setting operation
Description
.