Self Diagnosis On/Off Setting - Toshiba LF620 Instruction Manual

Electromagnetic flowmeter converter
Hide thumbs Also See for LF620:
Table of Contents

Advertisement

Switch operation
STEP1 
 
 
 
STEP2 
 
 
 
STEP3 
 
 
 
STEP4(=END) 
 
 
 

8.2.25 Self Diagnosis ON/OFF Setting

You can select on/off setting for the self-diagnosis function.
If the self-diagnosis function is set to OFF, no error message is displayed even if any of the errors listed
below occurs.
• ROM error
• RAM error
• System parameter error
• Excitation cable is not connected or its wiring is open
• Excitation circuit failure
• ADC circuit error
• Totalizer data is destroyed
OFF
ON
If this function is set to ON and an error occurs, an error message appears on the Display 2 measured
value screen. If an error occurs, the measurement items specified for Display 2 screen cannot be
displayed unless the error is removed.
Display example
EMPTY 
 
OFF 
 
 
 
ESC 
 
EMPTY 
 
OFF 
 
        
 
EMPTY 
 
SENS-HI 
 
        
 
EMPTY 
 
SENS-HI 
 
        
SET OK? 
ESC 
O K
Selection items
− 94 −
The current setting (OFF in this example)
appears.
Then push
* Pushing
ESC
screen.
The switches at the bottom change.
(
At the same time, the cursor appears.
(The item indicated by the cursor is
highlighted.)
Then push
You can continue to change the setting item.
Push
items.
: Selected item is scrolled up.
: Selected item is scrolled down.
When the desired item is selected, push
to set the item temporarily.
When you push
to confirm whether the setting is OK or not.
If OK, push
O K
setting, push
N O
Pushing
ESC
and exits the setting screen.
N O
Description
Self-diagnosis function is turned off.
Self-diagnosis function is turned on.
6 F 8 A 0 9 2 7  
Description
.
returns you to the menu
are shown.)
.
to change the selection
, a message appears
. If you want to redo the
.
cancels the setting operation