Converter Check / Calibration; Flow Rate Calibration (Zero Calibration) - Toshiba LF620 Instruction Manual

Electromagnetic flowmeter converter
Hide thumbs Also See for LF620:
Table of Contents

Advertisement

9.2 Converter Check / Calibration

9.2.1 0 % Flow Rate Calibration (Zero Calibration)

Using the converter's internal calibration circuit, 0% flow rate calibration (hereafter called zero
calibration) can be performed.
●Zero point check / calibration
Switch operation
STEP1 
 
 
STEP2 
 
 
Longer 
STEP3 
 
 
STEP4(=END) 
 
 
Note 1:
Note 2:
 
Display example
R:CAL 
1 CAL   0% 
2 CAL  50% 
3 CAL 100% 
 
4 EXC DSPL 
5 EXIT 
 
ESC
 
CAL   0% 
 
      0.1 
 
 
        % 
 
 
ESC 
 
ADJ READY 
 
      0.1 
 
 
        % 
        
 
O K
NOW 0% 
ADJUSTING 
 
          
O K
 
        
 
CAL   0% 
 
      0.0 
 
 
        % 
 
 
ESC 
 
To perform calibration, push and hold
To cancel the adjustment when ADJ READY is displayed, push
The screen returns to the zero display using the simulation input.
Select "CAL 0" from the setting item
selection menu.
 
When the calibration screen is selected, the
internal simulation circuit starts working and
0% value using the internal simulation
signal appears.
Then push and hold
* Pushing
ESC
menu.
The title of the screen changes to "ADJ
READY" and the converter is is ready for
calibration.
Pushing
N O
screen.
N O
Push
O K
rate.
 
It
takes
several
calibration for 0% flow rate and the
simulated value of 0% after calibration
appears.
Pushing
ESC
menu.
− 107 −
6 F 8 A 0 9 2 7  
Description
longer.
returns you to the setting
returns you to the previous
to start calibration for 0% flow
seconds
to
perform
returns you to the setting
longer.