7.4 Alarm and Shutdown messages

Alarm Message
Shutdown Message
Alarm/
Message
Shutdown
10VA of PEEP and Flow
Shutdown
Valve Power Failure
Apnea
User Alarm
Middle
Connect O
sensor?
User Alarm
2
Middle
CSB and Alarm MCU
Shutdown
Communication Failure
CSB and DMB
Shutdown
Communication Failure
DMB 3.3V voltage
Shutdown
Failure
Flow Sensor Reference
Shutdown
Voltage Failure
Inspiratory Flow Valve
Shutdown
(DAC) Failure
Inspiratory Flow Valve
Shutdown
(Drive) Failure
Low battery voltage
User Alarm
High
7-8
M1207027
Reproduced from the electronic master in MATRIX
This section covers the Alarm messages (not shaded) and Shutdown
messages (shaded) experienced by the ventilator system.
If a User Alarm persists after the recommended action has been
performed, the message indicates the probable component and
related circuit that needs repair. Use the Service Mode tests to isolate
the fault. The items in the Service Repair column indicate the path
from the named component to the control board.
Cause
Ventilator malfunction.
No mechanical breaths or
spontaneous breaths > 5 mL in
last 15 seconds.
The O
sensor is not
2
connected to the cable.
Ventilator malfunction.
Ventilator malfunction.
Ventilator malfunction.
Ventilator malfunction.
Ventilator malfunction.
Ventilator malfunction.
Voltage is <11V while using
battery power for 0.5 seconds.
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
User Action/ Concerns
Service Repair
Ventilate manually.
Replace:
Monitoring is not reliable.
• CSB
• DMB
• PEEP Valve
• Inspiratory Flow
Monitoring resumes after first
breath (mechanical) or 2
breaths within 30 sec
(nonmechanical).
Connect the O
sensor
If persists, check:
2
• O
• CSB
Ventilate manually.
Replace:
Monitoring is not reliable
• DMB
• CSB
• Cable from DMB
Ventilate manually.
Replace:
Monitoring is not reliable
• DMB
• CSB
• Cable from DMB
Ventilate manually.
Replace:
Monitoring is not reliable
• DMB
Ventilate manually.
Replace:
Monitoring is not reliable
• CSB
Ventilate manually.
Replace:
Monitoring is not reliable
• CSB
Ventilate manually.
Check/Replace:
Monitoring is still available.
• Flow Valve
• Flow Valve
• CSB
Manually ventilate the patient
to save power. Is the mains
indicator light on?
Make sure power is connected
and circuit breakers are
closed.
Valve
sensor cable
2
to CSB
to CSB
harness
M1207026