3.1 Ventilator post-service checkout
M1207026
M1207027
Reproduced from the electronic master in MATRIX
After servicing the 9100c ventilator, perform the following service
mode Calibrations:
•
O
Cell calibration
2
•
Paw Zero
•
Paw Gain
•
PEEP Valve
•
Flow Zero
•
Inspiratory Flow Valve
•
Flow Sensor Negative
•
Flow Sensor Positive
Then, complete the checkout procedure for the entire machine in the
following sections.