6.2 9100c Planned Maintenance
Serial Number:
Hospital:
12 months
24 months
6.2.1 Every twelve
(12) months
Parts Replacement
Machine Checks and
Tests
6-4
M1207027
Reproduced from the electronic master in MATRIX
Date: (YY/MM/DD)
Performed by:
_______________
This section covers the regular maintenance procedures (minimum
requirements) needed to make sure that the 9100c anesthesia
machine — including the ventilator — operates to specifications.
Perform the following steps every 12 months.
For details, refer to the sections listed.
Replace the vaporizer port o-rings (Section 9.10).
Refer to the 9100c User's Reference Manual.
Perform the following steps:
1. User maintenance listed below. Including disassembly,
inspection, cleaning and parts replacement as required.
•
Breathing Circuit Maintenance (Section 7.5)
•
Bellows Assembly Maintenance (Section 9)
•
Bellows Assembly Tests (Section 9)
•
O
Sensor Calibration (Section 9)
2
Refer to listed sections in this manual.
Perform the following steps:
1. MOPV pressure relief valve test (Section 6.5).
2. Exhalation valve reilef valve test (Section 6.6).
3. From the Service Mode, perform the following:
•
Display Discrete I/O Signals (Section 4.4.1)
Verify proper operations of all circuit boards and switches.
•
Adjust Drive Gas Regulator (Section 6.4)
•
PAW Zero (Section 4.5.1)
PAW Gain (Section 4.5.2)
•
•
PEEP Valve (Section 4.5.4)
•
Flow Zero (Section 4.5.5)
•
Inspiratory Flow Valve (Section 4.5.6)
Flow Sensor Positive (Section 4.5.7)
•
•
Flow Sensor Negative (Section 4.5.8)
4. Inspect the system (Section 3.2)
5. Pipeline and cylinder tests (Section 3.3)
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English