Self Tests - GE 9100c Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

4.1 Self tests

Powerup tests
Continuous tests
Periodic tests
M1207026
M1207027
Reproduced from the electronic master in MATRIX
The 9100c Ventilator software includes self tests that determine
whether or not the operating software is functioning properly and
whether or not the electronic circuits on the circuit boards are
functional.
The self tests include:
powerup tests
continuous tests
periodic tests
The following list of the tests run at powerup:
CSB and DMB communication
CSB and alarm CPU communication
Paw sensor reference voltage
Flow sensor reference voltage
DMB 3.3 V voltage
PEEP valve DAC and voltage feedback
Flow valve DAC and voltage feedback
10 VA of PEEP and flow valve
If one or more of these tests fail, the display provides a readout of the
problem.
These tests are run continuously during normal operation and alarms
are associated with each test. A failure causes an alarm to display on
the screen in the alarm display area.
Supply voltage checks
Battery voltage checks
These tests are run periodically during normal operation.
Tests that run every 1 second include:
DMB 3.3 V voltage
10 VA of PEEP and flow valve
Tests that run every 2 seconds include:
CSB and DMB communication
CSB and alarm MCU communication
If one or more of these tests fail, the display provides a readout of the
problem.
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
4 Service Tests and Service Mode