9100c
4.5.1 Paw Zero
Remarks
4-14
M1207027
Reproduced from the electronic master in MATRIX
The Paw Zero procedure
•
zeros any offset in the amplifier for the airway pressure
sensor.
It does so by reading the value for airway pressure when the
breathing system has been removed.
•
If the calibration passes, the calibration value can be stored
in the EEPROM.
•
If the calibration value is not within the correct tolerance, the
calibration fails.
Ensure that the breathing system is removed.
"Fail" indicates a problem in the CSB (Airway pressure transducer)
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English