4.5.1 Paw Zero

Remarks
4-14
M1207027
Reproduced from the electronic master in MATRIX
The Paw Zero procedure
zeros any offset in the amplifier for the airway pressure
sensor.
It does so by reading the value for airway pressure when the
breathing system has been removed.
If the calibration passes, the calibration value can be stored
in the EEPROM.
If the calibration value is not within the correct tolerance, the
calibration fails.
Ensure that the breathing system is removed.
"Fail" indicates a problem in the CSB (Airway pressure transducer)