4.6.4 Altitude Setting

Remarks
M1207026
M1207027
Reproduced from the electronic master in MATRIX
The Altitude Setting menu shows the altitude that is tailored to the
specific machine.
The accuracy of some of the ventilator measurements is altitude
sensitive. To ensure the specified accuracy, the altitude setting
should be set to the specific altitude where each machine is located.
Altitude settings range from -400 to 3600 meters in increments of 100
meters.
The setting is saved in EEPROM; the default value is 0 meter.
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
4 Service Tests and Service Mode
4-29