Hitachi L-2100 Instruction Manual page 83

Table of Contents

Advertisement

5.1
Checking the Performance and Specifications
5.1.2 Preparing the
System for
Operation
Load resistance coil
stainless pipe 0.1 mm i.d. × 10 m long(L-2100)
stainless pipe 0.25 mm i.d. × 10 m long(L-2130)
Waste liquid
bottle
(1)
Figure 5-1 shows how to connect the units for performance
check. According to this figure, carry out tubing.
Before energizing the system, ensure that the tubing
between the autosampler and pump/detector is properly
connected.
Detector (wavelength 250 nm, time constant 0.1 s)
Autosampler
Pump
Fig. 5-1
(2)
Connect the e-Line cable among the autosampler, detector
and pump.
(3)
Check if the drain tube is properly connected to the waste
liquid bottle. The waste liquid bottle must be at a lower
position than the autosampler.
5 - 2
Waste liquid
bottle
(Injection valve)
(Syringe valve)
Eluent Detergent
Flow Path Diagram