Battery Capacity Test (Ct) - Philips HEARTSTART XLT M3500B Instructions For Use Manual

Defibrillator/monitor
Table of Contents

Advertisement

M3500B HeartStart XLT Defibrillator/Monitor

Battery Capacity Test (CT)

To perform a Battery Capacity Test:
1. Turn the HeartStart XLT off.
2. Place a "Test in Progress" label on the HeartStart XLT to indicate to oth-
ers that it may not be used.
3. Insert a charged battery.
4. If an AC power module is connected, unplug the power module from the
HeartStart XLT. While pressing
5. Allow the test to proceed to completion. The test takes approximately
three hours and is complete when test results print out and the device
turns itself off.
6. Review the test results and take the appropriate action, as follows:
Table 11-1 Battery Capacity Test Results
If
Elapsed Time > 2.5 hours
and
Low Battery Time > 10 minutes
Elapsed Time < 2.5 hours
or
Low Battery Time < 10 minutes
Operational Checks
Mark
, press On to start the test.
Event
Then...
1.
The battery passed the test.
2.
Record "pass CT" and the date on
the bottom of the battery.
3.
Recharge the battery before use.
1.
The battery failed the test.
2.
Record "fail CT" and the date on
the bottom of the battery.
3.
Discard the battery appropriately.
11
11-7