ERC Protocol

NOTE
3-16
ERC Protocol
The HeartStart XLT can be configured to support the European Resuscitation
Council (ERC) Guidelines for Resuscitation (1998). If
figured to
, the defibrillation process described in this chapter is the same,
with the exception of how the Pause state functions (see "Pausing for CPR"
on 3-12).
As described, you can enter a Pause state:
at the completion of a shock series, or
when no shock is advised
When either of these events occurs, the ERC protocol prompts you to Check
Patient. Then it prompts you as follows:
Figure 3-14
If Needed
Press PAUSE
And Begin CPR
ANALYZE
PAUSE
Using the ERC protocol, you are not prompted to (or given time to) check the
patient's pulse.
Shocks: 1
00:00:50
l is con-
Defibrillating in AED Mode